FLUKE v3000 FC Wireless AC Voltage Kit FLUFLKV3000FC

Quantity
ADD TO CART
BUY IT NOW
  • Detail

    The Fluke v3000 FC WirelessVoltage Kit is part of the Fluke Connect team of wireless test tools

    Connect the v3000 FC Wireless AC Voltage Module to your test point, and view the results up to 20 meters away on the 3000 FC Wireless Digital Multimeter.

    You'll save time, with less run-around collecting multiple measurements. Use multiple modules for measurements on three-phase systems, as a stand-alone measurement tool or combine with other Fluke Connect® modules as a system for multiple measurements.

    From short distances, you can even view readings from modules through closed electrical panels. Plus no more writing down data as the FC Wireless Temperature Module captures up to 65,000 sets of time stamped min/max/avg readings, using the optional PC adapter. Fluke Connect wireless test tools also offer increased safety by letting you view readings in a separate location from the test point.

    Now you can take readings on moving machinery, with only the measurement module in harm's way.

    The Fluke FC Wireless v3000 Voltage Kit provides measurement of many common electrical parameters

    The Fluke 3000 FC Series Wireless Multimeter with the Fluke Connect® app has all the essentials for convenient test and measurement troubleshooting:

    • AC and DC voltage measurements to 1000 V
    • AC and DC current with 0.01 mA resolution
    • Continuity, resistance, diode test, capacitance and frequency measurements
    • MIN/MAX recording
    • CAT III 1000 V, Cat IV 600 V; IP54

    With the Fluke v3000 FC Wireless AC Voltage Module you can:

    • Measure up to 1000 V True-RMS AC
    • Use as a standalone meter or as part of the system
    • Use the logging function for recording and saving up to 65,000 readings

    Specifications:

    Fluke 3000 FC Wireless Digital Multimeter

    AC Voltage
    RangeResolutionAccuracyAccuracy
    45 Hz to 500 Hz500 Hz to 1 kHz
    600.0 mV0.1 mV1.0% + 32.0% + 3
    6.000 V0.001 V
    60.00 V0.01 V
    600.0 V0.1 V
    1000 V1 V

    DC Voltage, Continuity, Resistance, Diode Test and Capacitance
    FunctionRangeResolutionAccuracy
    mV600.0 mV0.1 mV0.09% + 2
    V6.000 V0.001 V0.09% + 2
    60.00 V0.01 V
    600.0 V0.1 V
    1000 V1 V0.15% + 2
    ��600 ��1 ��Meter beeps at < 25 ��, beeper detects opens or shorts of 250 ��s or longer.
    ��600.0 ��0.1 ��0.5% + 2
    6.000 k��0.001 k��0.5% + 1
    60.00 k��0.01 k��
    60.00 k��0.1 k��
    600.0 k��0.001 M��
    50.00 M��0.01 M��1.5% + 3
    Diode test2.000 V0.001 V1% + 2
    F1000 nF1 nF1.2% + 2
    10.00 ��F0.01 ��F
    100.0 ��F0.1 ��F
    9999 ��F1 ��F10% typical

    AC and DC Current
    FunctionRangeResolutionAccuracy
    mA AC
    (45 Hz to 1 kHz)
    60.00 mA0.01 mA1.5% + 3
    400.0 mA0.1 mA
    mA DC60.00 mA0.01 mA0.5% + 3

    Frequency
    RangeResolutionAccuracy
    99.99 Hz0.01 Hz0.1% + 1
    999.9 Hz0.1 Hz
    9.999 kHz0.001 kHz
    99.99 kHz0.01 kHz

    Input Characteristics
    FunctionOverload
    protection
    Input impedance
    (nominal)
    Common mode rejection ratio (1 k�� unbalance)Normal mode rejection
    1100 Vrms> 10 M�� < 100 pF> 120 dB at DC, 50 Hz or 60 Hz> 60 dB at 50 Hz or 60 Hz
    1100 Vrms> 10 M�� < 100 pF> 60 dB,DC to 60 Hz
    1100 Vrms> 10 M�� < 100 pF> 120 dB at DC, 50 Hz or 60 Hz> 60 dB at 50 Hz or 60 Hz
    Open circuit test voltageFull scale voltageTypical short circuit current
    To 6 M��50 M��
    1100 Vrms< 2.7 VDC< 0.7 VDC< 0.9 VDC< 350 mA
    1100 Vrms< 2.7 VDC2.000 VDC< 1.1 mA

    mA Function
    Overload ProtectionFused, 44/100 A, 1000 V FAST Fuse
    Overload600 mA overload for 2 minutes maximum, 10 minutes rest minimum

    MIN/MAX Recording Accuracy
    DC functionsThe specified accuracy of the measurement function ��12 counts for changes > 350 mS in duration.
    AC functionsThe specified accuracy of the measurement function ��40 counts for changes > 900 mS in duration.

    General Specifications
    Maximum voltage between any terminal and earth ground1000 VDC or AC rms
    �� fuse protection from A inputs0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse, Fluke specified part only
    Display (LCD)Update rate4/sec
    Volts, amps, ohms6000 counts
    Frequency10,000 counts
    Capacitance1,000 counts
    Battery typeThree AA Alkaline batteries, NEDA 15A IEC LR6
    Battery life250 hours minimum
    RF communications2.4 GHZ ISM Band
    RF communication rangeOpen air, unobstructedUp to 20 m
    Obstructed, sheetrock wallUp to 6.5 m
    Obstructed, concrete wall, or steel electrical enclosureUp to 3.5 m
    TemperatureOperating-10°C to 50°C
    Storage-40°C to 60°C
    Temperature coefficient0.1 x (specified accuracy) / °C (< 18°C or > 28°C)
    Relative humidity0% to 90% (0°C to 35°C)
    0% to 75% (35°C to 40°C)
    0% to 45% (40°C to 50°C)
    AltitudeOperating2,000 m
    Storage12,000 m
    Electromagnetic compatibility
    EMI, RFI, EMC, RF
    EN 61326-1:2006, EN 61326-2-2:2006
    ETSI EN 300 328 V1.7.1:2006, ETSI EN 300 489
    V1.8.1:2008, FCC Part 15 Subpart C Sections 15.207, 15.209, 15.249
    FCCID : FCC: T68-FDMMBLE IC: 6627A-FDMMBLE
    Safety complianceANSI/ISA 61010-1 / (82.02.01)3rd edition
    CAN/CSA-C22.2 No 61010-1-123rd edition
    UL 61010-13rd edition
    IEC/EN 61010-12010
    CertificationsCSA, FCC, CE
    Ingress Protection (IP) ratingIP54
    Pollution degree2

    Fluke v3000 FC Wireless AC Voltage Module

    Specifications
    Range / Resolution6.000 V / 0.001 V
    60.00 V / 0.01 V
    600.0 V / 0.1 V
    1000 V / 1 V
    Accuracy45 Hz to 500 Hz1.0% + 3
    500 Hz to 1 kHz2.0% + 3

    General Specifications
    LCD w/ backlight3 digits, 6000 counts, updates 4/sec
    Battery type2 AA, NEDA 15 A, IEC LR6
    Battery life400 Hours
    MemoryRecord up to 65000 readings
    RF communications2.4 GHZ ISM Band
    RF communication rangeOpen air, unobstructedUp to 20 m
    Obstructed, sheetrock wallUp to 6.5 m
    Obstructed, concrete wall or steel electrical enclosureUp to 3.5 m
    Operating temperature-10°C to +50°C
    Storage temperature-40°C to +60°C
    Temperature coefficient0.1 x (specified accuracy) / °C (< 18°C or > 28°C)
    Operating humidity90% at 35°C, 45% at 40°C, 45% at 50°C
    AltitudeOperating2,000 m
    Storage12,000 m
    EMCEN 61326-1:2006
    Safety complianceANSI/ISA 61010-1 / (82.02.01)3rd edition
    CAN/CSA-C22.2 No 61010-1-123rd edition
    UL 61010-13rd edition
    IEC/EN 61010-12010
    Safety ratingCAT IV 600 V, CAT III 1000 V
    CertificationsCSA, CE, FCC: T68-FBLE IC: 6627A-FBLE
    Ingress Protection (IP) ratingIP42
    Pollution Degree2